Xenon aging test chamberIt has become a common choice in the field of light resistance testing, providing sufficient technical reference and practical proof for related industries. This instrument is a test instrument that uses a long arc xenon lamp as a light source to simulate and enhance weather resistance and accelerate aging. It is used to quickly obtain near atmospheric aging test results and evaluate the weather resistance of materials such as plastic rubber, paint coatings, textile printing and dyeing. The instrument meets relevant standards such as GB/T16422-2, ASTMD529, ISO4892-2, etc | |
Technical Parameter | |
power supply |
AC 380V/50Hz, maximum power 9KW |
Xenon lamp power |
3KW (air-cooled full solar spectral long arc xenon lamp) |
Xenon lamp radiation intensity |
Adjustable setting value (350-850nm) 550W/m2 |
Trigger power |
6KW working voltage AC 220V current 27 ± 3A, |
Temperature range |
63 ± 5 ℃, test blackboard temperature: 40-110 ℃ |
temperature uniformity |
± 2℃ |
Rotation speed of the sample rotating basket |
≥ 1r/min (stepless speed regulation) |
Studio Size |
950×950×950 |
filter |
0% below 255nm, 90% above 400-800nm. (3-6 months replacement) |
Product Features | |
1. Ventilation and cooling device: The temperature ventilation and cooling system of the test chamber is automatically controlled by a temperature controller | |
2. The control blackboard thermometer adopts wireless transmission technology, effectively improving the control accuracy. | |
3. Use fiber optic transmission to measure the radiation intensity of xenon lamps, avoiding measurement errors caused by temperature. | |
4. The sample holder rotates around the light source at a speed of 1r/min, and a blackboard thermometer BPT or a black standard thermometer BST is installed on the sample holder | |
5. Timing device: In order to record the working time of each experiment, the instrument is equipped with a timer on the operation panel, which can control the instrument time. When the timer ends, the instrument will automatically shut down. | |
6. Protection device: In order to ensure the normal operation of the xenon arc lamp and prevent damage, this instrument is equipped with a protective device that, in addition to the xenon arc lamp being ignited in the circuit and interlocked with the cooling water (i.e. the xenon arc lamp cannot be lit when the cooling water is not working), automatically extinguishes the xenon lamp and stops the entire machine from working if there is a cooling water malfunction during operation. | |