brand | Other brands | Price range | 5000-10000 |
---|---|---|---|
Instrument type | laboratory | Flow range | Small traffic |
Origin category | domestic |
Y09-3016 Handheld Air Dust Particle Counter
Y09-3016 Handheld Air Dust Particle CounterThe company has successfully developed a full half body laser sensorHandheld laser dust particle counterIt can be connected to a PC computer data acquisition system for remote control, allowing for direct observation of instrument testing conditions. Test data can be analyzed and processed through a computer and saved as Excel files. The whole machine adopts American microcomputer control processing technology, semiconductor laser sensor technology, and imported air pump technology, which has the characteristics of multiple functions, high measurement accuracy, fast speed, easy portability, and simple operation. The instrument can simultaneously measure the number of dust particles of multiple sizes in one sampling, and this product has been widely used in clean room testing of electronic production enterprises; On site inspection and leak detection of filters; It can monitor biosafety, HVAC systems, computer rooms, beverage packaging environments, drug and medical device production environments, hospital clean operating rooms, automotive spraying environments, microelectronics, biochemical products, food hygiene, fine chemicals, precision machinery and other production and research departments. It is the instrument selected by HVAC and pharmaceutical enterprises and their supervision and management departments to implement GMP standards and electronic production enterprises.
Main technical parameters
light source |
Semiconductor laser with a lifespan of over 35000 hours |
Sampling traffic |
2.83L/min (0.1ft3) imported sampling pump |
Display mode |
LCD display, real-time display, previous cycle display, real-time concentration display, can display time, date, measurement value, temperature and humidity, room number, sampling point, sampling frequency, battery level, status and other parameters, 95% UCL calculation, can directly display particle concentration (particles/cubic meter) |
Rechargeable battery |
Lithium ion battery, 7.4 V, 2200mAh |
Power supply |
AC power adapter, AC :100V~245 V,50/60 HzDC:7.4 V,1 A; |
working hours |
6 hours (remaining battery indicator) |
Counting mode |
Cumulative value, difference value, concentration value |
Testing method |
Single, repetitive, continuous, computational, remote |
Unit conversion |
The unit can be converted into m3 |
Particle size channel |
Counting six particle sizes simultaneously: 0.3, 0.5, 1.0, 3.0, 5.0, and 10 μ m |
sampling period |
1~10000 (s) Delay count: 0~99 (s) Self purification time: ≤ 10 (min ) |
work environment |
Temperature: 10-40 ℃ (50-104 ℉) Relative humidity: 20-90% RH, no condensation Atmospheric pressure: 86-106KPa |
Temperature and humidity: |
Choose (1) Temperature: 0-50 ℃± 1 ℃ (2) Humidity: 0-100% RH ± 5% |
UCL settings |
Sampling points (A): Set from 2 to 9 points. Sampling frequency per point: (L) 2-9 times Set measurement position: 0-999 |
UCL report |
Compliant with ISO14644-1, GB50073-2001, GMP, FS-209E standards |
data storage |
It can store 1000 sets of data (circular buffer) (including particle size, data, environmental data, year, month, day, time, sampling volume, data location port), and the data will not be lost after power failure. |
communication interface |
RS232, 9600 baud rate |
Alarm settings |
The instrument is equipped with an alarm function, which can alert the cleanroom of levels 100, 1000, 10000, 100000, 300000, and 1000000 when the battery voltage exceeds the standard |
Zero Count |
Meets the requirements of JJG-547-88 Dust Particle Counter Verification Regulations JIS B9921:1997, with less than one count every five minutes |
Overlap error |
At 5%, 2000000 particles per cubic foot |
Printing function |
External printer (optional) |
External dimensions |
230×130×45(mm) |
weight |
0.6kg |
calibration |
Traceable to the National Institute of Standards and Technology (NIST) in the United States, our company has passed the metrological standard building assessment. Traceable to the Shanghai Institute of Metrology and Measurement Technology, we can also calibrate ourselves or third-party metrology institutions |
Standard configuration |
Suitcase/charger, equal power sampling head, zero filter, RS 232 connection cable, computer connection communication software CD |
Optional parts |
External printer, professional triangular sampling rack, temperature and humidity sensor |