0.05% power measurement accuracy
PA2000mini is the only power analyzer in the industry that can perform mobile and portable measurements. While further optimizing the volume and weight, the power measurement accuracy can also ensure 0.05%, making it the best testing partner for outdoor photovoltaic on-site acceptance testing and high-speed rail inverter testing.
On site acceptance of photovoltaic wind power
Many solar photovoltaic power plants and wind power plants are located in remote mountainous areas or on rooftops, with harsh conditions that often make power supply difficult. This requires power analyzers to be more portable while testing accurately. The PA2000mini power analyzer has a small size, comes with a battery, and can work continuously for 3-4 hours, making it very suitable for on-site acceptance environments of photovoltaic and wind power
High speed railway inverter detection
With the rapid development of electric locomotives, online detection of inverters has become an inevitable trend. The ZIT6000 portable inverter detector, with its compact design and rigorous testing, can come and go freely even in small distribution boxes. We have achieved high-precision and high bandwidth analysis for mobile testing, fully meeting your testing needs in all aspects.
system integration
The PA2000mini power analyzer is dedicated to providing you with a convenient testing experience. With a 4-channel 0.05% precision power measurement board and a motor testing board, most daily measurement applications can be easily completed. The standard multiple protocol interfaces, combined with the rack kit, make system integration more convenient.
PA2000mini power analyzer industry application
●Testing and Certification Laboratory
●Photovoltaic and wind power industry
●Inverter and motor industry
●Electric vehicle industry
●Charging pile industry
●Robot industry
●Power industry
●Transformer industry
Technical parameters of PA2000mini power analyzer
input parameter | parameter description | ||
Measurement range | voltage | 1.5V, 3V, 6V, 10V, 15V, 30V, 60V, 100V, 150V, 300V, 600V, 1000V, 1500V (peak factor 1.33) | |
electric current | Direct input (peak factor of 3) | ||
5A: 10mA, 20mA, 50mA, 100mA, 200mA, 500mA, 1A, 2A, 5A | |||
Sensor input (peak factor of 3) | 50mV, 100mV, 200mV, 500mV, 1V, 2V, 5V, 10V | ||
Continuous maximum allowable input value | voltage | Peak value of 2000V or RMS value of 1300V, whichever is smaller | |
electric current | Direct Input | ||
5A: Peak value of 15A or RMS value of 6.5A, whichever is smaller | |||
Sensor Input | The peak value shall not exceed 4 times the range, and the effective value shall not exceed 2 times the range | ||
Instantaneous maximum allowable input value (≤ 1s) |
voltage | Peak value of 3000V or RMS value of 1500V, whichever is smaller | |
electric current | Direct Input | ||
5A: Peak value 22.5A or RMS value 10A, whichever is smaller | |||
Sensor Input | The peak value shall not exceed 5 times the range, and the effective value shall not exceed 3 times the range |
Input impedance | PA2000mini |
Voltage input impedance | Input resistance: 2M Ω, input capacitance: 10pF |
Current input impedance | Input resistance: 100m Ω, input inductance: 0.07 μ H |
Sensor voltage input impedance | Input resistance: 1M Ω, input capacitance: 45pF |
input parameter | PA2000mini |
Input Bandwidth | DC,0.1Hz~500KHz |
sampling rate | 500KS/s |
Maximum continuous common mode voltage | 1000Vrms |
Accuracy:
Input signal frequency range | PA2000mini | |
Voltage/Current | power | |
DC | Direct current input: 0.05+0.10+20uA Voltage input: 0.05+0.10 Sensor: 0.05+0.05 |
Direct current input: 0.05+0.10+20uA * voltage reading Voltage input: 0.05+0.10 Sensor: 0.05+0.10 |
0.1Hz ≤ f < 30Hz | 0.10 + 0.20 | 0.20 + 0.40 |
30Hz ≤ f < 45Hz | 0.10 + 0.10 | 0.10 + 0.20 |
45Hz ≤ f < 66Hz | 0.05 + 0.05 | 0.05 + 0.05 |
66Hz ≤ f < 1KHz | 0.10 + 0.10 | 0.20 + 0.10 |
1KHz ≤ f < 10KHz | 0.20 + 0.10 | 0.30 + 0.20 |
10KHz ≤ f < 50KHz | 0.30 + 0.10 | 0.30 + 0.20 |
50KHz ≤ f < 100KHz | 2.00 + 0.50 | 2.00 + 1.00 |
100KHz ≤ f < 500KHz | 5.00 + 1.00 | 8.00 + 2.00 |
500KHz ≤ f ≤ 1MHz | 6.00 + 1.00 | 10.00 + 2.00 |
The measurement accuracy of the power analyzer is given under the following conditions: temperature: 23 ± 1 ℃; Humidity: 30~70% R.H; Input signal: sine wave; Common mode voltage: 0V; after preheating, line filter: OFF。 λ (power factor): 1. Peak factor: 3. After preheating for 30 minutes, f is the frequency. Data update rate: 500ms.
Display:
Display parameters | PA2000mini |
monitor | 9 "color LCD display |
resolution | 800 × 480 pixels |
Touch Screen | Support touch screen operation |
Display updates | Same as the data update rate |